Optimize your battery when you first receive it:
- Place the rechargeable battery in the oximeter
- Charge fully (4 hours) then use oximeter
- Do not continually charge the battery when in use as the battery
loses performance. It is better to run on battery and then recharge
when not in use
- If reliable electricity is available, the oximeter may be used
continually plugged in, but it is better to remove the battery in this
situation
Thereafter use/recharge as required. The oximeter should run for 12-16
hours between charges.
Readings may become inaccurate when the battery is
very low, so recharge in time!